3D MRI Geometric Distortion, Synthetic-CT Physics Audit, Electron-Return-Effect (ERE) & Formal IEC 60601-2-33 RF/Implant Safety Constraint Engine — with Closed-Loop Sequence Co-Design Agent & MCP Server

Python 3.11 | 3.12 License: Apache-2.0 Tests: 33 Passed Coverage: 92.6% Ruff: 0 Errors MCP: JSON-RPC 2.0 Research Software

RESEARCH SOFTWARE — NOT A MEDICAL DEVICE. mrphys-guard is open-source computational medical physics software for research, algorithm verification, and education. It is not cleared or approved by the U.S. FDA, EU MDR, or any regulatory authority, and must never be used for clinical patient clearance, implant MR-safety sign-off, or radiation therapy treatment planning. Generic implant envelopes included in this repository are literature reference models and never replace manufacturer Instructions For Use (IFU) or qualified Medical Physicist / MR Safety Officer (MRSO) review.


Why mrphys-guard?

Consumer AI chat interfaces cannot solve 3D magnetostatic Maxwell convolutions, reconstruct spherical-harmonic gradient-coil distortion tensors across a 400 mm DSV, detect silent bone-to-air swaps in AI synthetic CTs, evaluate Lorentz-force electron-return-effect (ERE) boundary spikes (+20–40% at 1.5T), or run formal IEC 60601-2-33 / ASTM F2503 implant constraint satisfaction with hard safety abstentions.

mrphys-guard provides an exact, deterministically verifiable MR-Physics Core Engine, a Closed-Loop Sequence & Safety Co-Design Agent gated by an independent Critic-Verifier, and a JSON-RPC 2.0 stdio MCP Server backed by a SHA-256 cryptographic provenance ledger.

Related MCP server: slicer-mcp

System Architecture

flowchart LR
    IN["Clinical Inputs<br/>DICOM MR/CT • NIfTI • NRRD<br/>3D Grid Phantom • Implant Spec"] --> IO["DICOM / NIfTI / NRRD I/O<br/>(dicom_io.py + ir.py)"]
    IO --> GNL["3D Solid Spherical Harmonics<br/>l ≤ 5 System GNL Solver<br/>(distortion_solver.py)"]
    IO --> B0["3D k-Space Lorentz-Sphere<br/>Susceptibility Dipole Solver<br/>ΔB0(k) = B0(1/3 − kz²/|k|²)χ(k)"]
    IO --> SCT["sCT Plausibility Auditor<br/>& 1.5T/0.35T ERE Kernel<br/>(sct_ere_auditor.py)"]
    IO --> CSP["Formal IEC 60601-2-33 /<br/>ASTM F2503 Implant CSP<br/>(implant_safety_csp.py)"]

    GNL --> AGENT["Closed-Loop MR Sequence<br/>Co-Design Agent<br/>(agent_orchestrator.py)"]
    B0 --> AGENT
    SCT --> SOBOL["Saltelli–Jansen QMC<br/>Global Sobol' Sensitivity<br/>(sobol_sensitivity.py)"]
    CSP --> GATE{"Hard Safety<br/>Abstention Gates<br/>(provenance.py)"}

    GATE -- "MR_UNSAFE / UNSPECIFIED" --> ABSTAIN["ABSTAIN + Ledger"]
    GATE -- "MR_CONDITIONAL" --> AGENT
    AGENT --> CRITIC["Independent Critic-Verifier<br/>Gate (APPROVED / REJECTED)"]
    CRITIC --> OUT["AAPM TG-284 / IEC 60601-2-33<br/>Dossier + SHA-256 Ledger<br/>& 6-Tool MCP Server"]

Verified Benchmark Dashboard

All numbers below are reproducible via uv run python scripts/run_benchmarks.py and logged in results/benchmark_summary.json:

Verification Benchmark

Metric

Verified Result

Reference / Gate

Lorentz-Sphere Dipole Identity

Interior mean $\Delta B_0$ / RMS

0.000 ppm / 0.019 ppm

Exact analytical $0\text{ ppm}$ (out of 9.41 ppm)

Lorentz-Sphere Exterior Decay

Relative $L_2$ error vs. $\frac{1}{3}B_0\chi_0(R/r)^3(3\cos^2\theta-1)$

0.54% (0.0054)

Marques & Bowtell (2005)

Bandwidth Scaling Law

$\Delta r_{B_0} \propto 1/\mathrm{BW}_{\mathrm{px}}$

Exact (rtol = 1e-14)

IEEE-754 machine precision

Maxwell Spherical Harmonics QA

$\ell \le 3$ vs. $\ell \le 5$ residual RMS ($\sigma = 0.02\text{ mm}$)

0.217 mm $\to$ 0.034 mm

AAPM TG-284 (2021); noiseless < 1e-6 mm

ERE Field Dependence (Real 3D Slicer Brain CT)

Peak exit spike (0T / 0.35T / 1.5T / 3.0T)

0.0% / +14.6% / +43.2% / +53.6%

Strictly 1.5T > 0.35T > 0T (Raaijmakers 2005/2008)

ERE Field Dependence (Pelvic SRS Phantom)

Peak exit spike (0T / 0.35T / 1.5T / 3.0T)

0.0% / +5.6% / +28.7% / +42.6%

Relativistic $r_g = 4.33\text{ mm}$ (1.5T) vs. 18.54 mm (0.35T)

Agent Co-Design: Prostate SRS + Ti Hips + Gas

Target $

\Delta\mathbf{r}

$ & CSP Verdict

Agent Co-Design: Pacemaker Cardiac + Brain

Peak $B_{1+\text{rms}}$ / WB SAR

3.87 uT / 2.93 W/kg $\to$ 1.88 uT / 0.69 W/kg

Critic APPROVED (<= 2.0 uT, <= 2.0 W/kg)

Hard Safety Abstention Gate

Ferromagnetic Aneurysm Clip (MR_UNSAFE)

ABSTAIN (MR_UNSAFE_DEVICE_DETECTED)

Zero bypass permitted

Saltelli–Jansen Global Sobol' ($N=1024$)

Dominant variance drivers (Geom / Dose)

BW (0.46) & GNL (0.43) / Δχ-ERE (0.45) & sCT (0.43)

Saltelli et al. (2010), 6,144 evals/output

Lossless DICOM Round-Trip

CT HU error / MR pixel error

0.0 HU / 5.2e-6

Real pydicom headers preserved

3D DCT Phase Unwrapping (v0.2.0)

Multi-$2\pi$ cycle recovery error

8.88e-16 rad (machine precision)

Ghiglia & Romero (1994) 3D Poisson DCT

Distortion Jacobian Audit (v0.2.0)

Low-BW folding singularity detection

min J = -10.12 (2,520 folding vox)

Intensity-preserving unwarping & pile-up alert

Multi-Beam 7-Field ERE (v0.2.0)

Single-beam vs. 7-beam IMRT net peak

+31.5% $\to$ +8.2% (-73.9% reduction)

Opposing-beam cancellation & gantry avoidance

Multi-Implant CSP + IEC PNS (v0.2.0)

Pacemaker + Ti Hip co-existence & PNS

PASS (pacemaker governs 2.0 uT)

UNSAT conflict cores + IEC 80%/100% PNS

Bloch CNR Pareto Frontier (v0.2.0)

3 Critic-APPROVED clinical protocols

0.394 mm / 0.478 mm (CNR 1.89) / 2.28 uT (CNR 2.67)

Distortion vs. Bloch CNR vs. RF/SAR trade-off


Publication-Grade Visuals (300-DPI PNG + PDF)

All six figures are generated reproducibly by scripts/generate_figures.py (Okabe–Ito colorblind-safe palette, explicit mm scalebars, and provenance footers) and visually audited in docs/AUDIT_REPORT.md.

1. 3D Geometric Distortion Decomposition (System GNL + k-Space Susceptibility Dipole)

Figure 1: 3D MRI Geometric Distortion Decomposition

2. Receiver Bandwidth Tradeoff & Multi-Field Distortion–SNR Pareto Frontier

Figure 2: Bandwidth vs Distortion and SNR Pareto Frontier

3. Synthetic-CT Physics Plausibility Audit & 1.5T vs. 0.35T Lorentz-Force ERE Hotspots

Figure 3: Synthetic-CT Audit and ERE Boundary Risk

4. Formal IEC 60601-2-33 / ASTM F2503 Implant Safety CSP Envelope (Before vs. Co-Designed)

Figure 4: Implant Safety CSP Envelope

5. AAPM TG-284 3D Grid-Phantom Solid Spherical-Harmonic QA ($\ell \le 5$)

Figure 5: 3D Grid-Phantom Spherical Harmonics QA

6. Saltelli–Jansen QMC Global Sobol' Sensitivity Analysis ($S_1$ vs. $S_T$)

Figure 6: Saltelli-Jansen QMC Global Sobol Sensitivity

7. Multi-Beam IMRT ERE Cancellation & 3D DCT Dual-Echo Phase Unwrapping (v0.2.0)

Figure 7: Multi-Beam ERE Cancellation and 3D DCT Phase Unwrapping


Workflows (Quickstart, Python API & MCP Server)

# 1. Create Python 3.12 virtual environment and install dependencies
uv sync --python 3.12 --extra dev

# 2. Fetch SHA-256-verified public open-license paired CT-MR Brain data (3D Slicer)
uv run python scripts/fetch_public_mri_sample.py

# 3. Run full test suite with coverage and ruff lint
uv run ruff check src tests scripts
uv run pytest --cov=src/mrphys_guard --cov-report=term-missing

# 4. Run end-to-end benchmarks and regenerate all 6 publication figures
uv run python scripts/run_benchmarks.py
uv run python scripts/generate_figures.py

# 5. CLI subcommands
uv run mrphys-guard list-implants
uv run mrphys-guard codesign --preset prostate_srs_hip_gas
uv run mrphys-guard codesign --preset pacemaker_cardiac_brain
uv run mrphys-guard sobol --n-base 1024
from mrphys_guard.distortion_solver import (
    build_pelvic_prostate_susceptibility_phantom,
    compute_3d_distortion_decomposition,
    fit_spherical_harmonic_gnl,
    generate_synthetic_grid_phantom,
)
from mrphys_guard.harness.agent_orchestrator import (
    MRSequenceCoDesignAgent,
    build_prostate_srs_hip_gas_scenario,
)
from mrphys_guard.ir import MRSequenceSpec
from mrphys_guard.sct_ere_auditor import compute_ere_boundary_risk

# 1. Fit l <= 5 Maxwell solid spherical harmonics on a 3D grid phantom
true_pts, meas_pts, _ = generate_synthetic_grid_phantom(noise_sigma_mm=0.02)
gnl_model = fit_spherical_harmonic_gnl(true_pts, meas_pts, lmax=5)

# 2. Solve 3D k-space susceptibility dipole convolution + GNL decomposition
phantom = build_pelvic_prostate_susceptibility_phantom(shape=(64, 64, 48), voxel_size_mm=(2.5, 2.5, 2.5))
seq = MRSequenceSpec(name="Pelvis_GRE", b0_T=1.5, bandwidth_hz_per_px=500.0, voxel_size_mm=(2.5, 2.5, 2.5), tr_ms=15.0, te_ms=4.5)
summary, fields = compute_3d_distortion_decomposition(
    chi_si=phantom["chi_si"],
    sequence=seq,
    gnl_model=gnl_model,
    target_roi_mask=phantom["prostate_mask"],
    body_mask=phantom["body_mask"],
)

# 3. Evaluate 1.5T MR-Linac Lorentz-force ERE boundary hotspot risk
ere_risk, ere_maps = compute_ere_boundary_risk(phantom["ct_hu"], (2.5, 2.5, 2.5), b0_T=1.5)

# 4. Run closed-loop MR sequence & implant safety co-design agent
result = MRSequenceCoDesignAgent().run(build_prostate_srs_hip_gas_scenario())
print(result.status, result.critic.status, result.optimized_max_target_distortion_mm)

Start the stdio MCP server with:

uv run mrphys-guard-mcp

Add to your MCP client configuration:

{
  "mcpServers": {
    "mrphys-guard": {
      "command": "uv",
      "args": ["run", "--directory", "/path/to/mrphys-guard", "mrphys-guard-mcp"]
    }
  }
}

Exposed provenance-gated tools:

  1. mrphys_solve_3d_distortion — 3D GNL ($\ell \le 5$) + k-space Lorentz-sphere susceptibility dipole solver

  2. mrphys_audit_sct_and_ere — Synthetic-CT physics-plausibility auditor + 1.5T/0.35T ERE boundary risk kernel

  3. mrphys_verify_implant_csp — Formal IEC 60601-2-33 / ASTM F2503 implant safety constraint verifier (incl. IEC PNS %)

  4. mrphys_codesign_safe_sequence — Closed-loop sequence & safety co-design agent with independent Critic-Verifier gate

  5. mrphys_lookup_implant_profile — Built-in generic MR-conditional / MR-unsafe device profile registry

  6. mrphys_run_sobol_sensitivity — Exact Saltelli–Jansen QMC global Sobol' sensitivity analysis ($S_1, S_T$)

  7. mrphys_multibeam_ere_optimize — Multi-beam IMRT/VMAT opposing-beam ERE cancellation & low-ERE gantry ranking

  8. mrphys_verify_multi_implant_csp — Simultaneous multi-implant CSP with governing bottlenecks & UNSAT conflict cores


Documentation, Governance & Citations

Available Tools

8 tools
mrphys_audit_sct_and_ereC

Audit synthetic-CT (sCT) physics plausibility (bone-to-air swaps, missing metal, body-contour truncation) and compute 1.5T/0.35T Lorentz-force ERE boundary hotspot risk.

Inputs: b0_T, grid_size, beam_energy_mv, swap_bone_fraction, +1 more.

mrphys_codesign_safe_sequenceB

Run the closed-loop MR sequence & safety co-design agent with independent Critic-Verifier gate on a preset or custom high-risk clinical scenario.

Inputs: preset, custom_scenario.

mrphys_lookup_implant_profileA

Lookup built-in generic ASTM F2503 / IEC 60601-2-33 MR-conditional literature device profiles (pass device_id or omit to list all profiles).

Inputs: device_id.

mrphys_multibeam_ere_optimizeC

Evaluate multi-beam IMRT/VMAT opposing-beam Lorentz-force ERE cancellation and recommend lowest-ERE gantry angles at 1.5T or 0.35T.

Inputs: b0_T, grid_size, gantry_angles_deg.

mrphys_run_sobol_sensitivityB

Run exact Saltelli-Jansen QMC global Sobol' sensitivity analysis (S_1, S_T) over receiver bandwidth, air-cavity susceptibility, GNL residual, and sCT HU error.

Inputs: b0_T, seed, n_base.

mrphys_solve_3d_distortionC

Solve 3D MRI gradient nonlinearity (l<=5 solid spherical harmonics) and k-space Lorentz-sphere susceptibility B0 dipole convolution on a pelvic prostate MR-only phantom.

Inputs: b0_T, grid_size, srs_tolerance_mm, include_rectal_gas, +2 more.

mrphys_verify_implant_cspB

Formally verify a multi-sequence MRI protocol against IEC 60601-2-33 / ASTM F2503 implant constraints (B0, dB/dx, B1+rms, SAR, slew rate, dB/dt, active scan time).

Inputs: sequences (required), implant_id (required), scanner, implant_override.

mrphys_verify_multi_implant_cspC

Formally verify a multi-sequence MR protocol simultaneously across multiple co-existing implants, returning intersected limits, IEC PNS %, and UNSAT conflict cores.

Inputs: sequences (required), implant_ids (required), scanner.

Tool Schema Changelog

Recent tool additions, removals, and schema changes observed during successful MCP inspections.

Loading tool schema changelog…

    TDQS

    B3.3/5.0

    Scored across 8 tools

    Disambiguation4/5

    Each tool maps to a fairly distinct phase of MR physics/safety workflow (solve, audit, verify, co-design, lookup, sensitivity, optimize). The clearest overlap is mrphys_verify_implant_csp vs mrphys_verify_multi_implant_csp, and mrphys_audit_sct_and_ere shares ERE territory with mrphys_multibeam_ere_optimize, but descriptions clarify single-vs-multi-implant and audit-vs-optimize distinctions.

    Naming Consistency5/5

    All eight tools use the mrphys_ prefix with a consistent snake_case verb_noun pattern (solve_, audit_, verify_, codesign_, lookup_, run_, multibeam_...). Compound nouns like verify_multi_implant_csp are long but follow the same convention throughout.

    Tool Count5/5

    Eight tools is well-scoped for a specialized MR-physics safety verification server; each tool covers a distinct analysis stage with no filler. Nothing feels thin or bloated for the stated domain.

    Completeness4/5

    The surface covers simulation, audit, single/multi-implant CSP verification, co-design, sensitivity analysis, beam optimization, and implant profile lookup—a coherent safety lifecycle. Minor gaps: no ability to register/edit custom implant profiles (lookup is read-only over built-ins) and no explicit report/export or phantom-definition tool.

    Related MCP Connectors

    Related MCP Servers

    • A
      license
      Not graded
      quality
      B
      maintenance
      Enables vision-language-model agents to examine clinical CT volumes the way a clinician does — orbiting and zooming the 3D view, applying volume-rendering presets, clipping, scrolling slices, measuring HU and distances, and showing or isolating segmented structures — using only named, range-checked tools with no code execution. Every call is logged with its arguments, resulting state and rendered frames so runs can be audited, replayed and scored as training data.
      Apache 2.0
    • F
      license
      B
      quality
      C
      maintenance
      Exposes five typed, provenance-gated physics tools over JSON-RPC 2.0 stdio so frontier agents can compute 3D deformable biophysical dose accumulation, DIR-uncertainty envelopes, and inverse remaining-dose ceilings for multi-course re-irradiation. Every output is audited by a critic-verifier gate with a hard abstention check, returning safety-flagged dossiers and provenance ledgers instead of unsupported clinical numbers.
      7
      -
    • F
      license
      B
      quality
      C
      maintenance
      Enables agents to run provenance-gated medical-physics computations over JSON-RPC 2.0 stdio, including TG-233 image-quality metrology, dose-detectability Pareto optimization, peak-skin-dose analysis, and audit workflows.
      10
      1
      -