Skip to main content
Glama

ai-bench-psu

Experimental. Use at your own risk. This software drives a laboratory power supply that can deliver 60 V or 20 A, enough to start fires, destroy equipment and injure people. Read the code yourself before you point it at real hardware. A software bug, a wrong setting, an agent misreading a tool, a sensor reading wrong, or a firmware quirk can energise the output or change its voltage or current. You are entirely responsible for any damage, injury, data loss or downtime that results from using it. The authors and contributors accept no liability.

Headless control of one bench power supply and the temperature sensors next to it, from a small Linux host, for people and for AI agents. Everything goes through drivers chosen in the config: one PSU driver per daemon, plus any number of sensor drivers (a device may carry several sensors).

  • Daemon (bench-psud): the only process that owns the supply's port and the sensors. Polls the supply at 4 Hz, reads each sensor at its own interval, serves a JSON API, a status page and an MCP endpoint, keeps Wh / Ah counters, logs every state change and a periodic sample as JSON lines, and runs the opt-in temperature interlocks, which switch the output off (never on).

  • CLI (benchctl): talks to the daemon; --direct opens the supply when the daemon is stopped. kiprim is the same program under its earlier name.

  • Live display (benchtop, optional service): big volts and amps, a V/A chart that goes to a fine scale when the values barely move, a temperature chart with the interlock limits, on a terminal or a monitor plugged into the host. It reads the daemon's API only.

  • Drivers: owon-scpi (OWON SPE/SPS-family SCPI over USB serial; tested on a Kiprim DC620S), temper (PCsensor-type TEMPer USB thermometers over hidraw), and the simulated sim and sim-temp used by the tests and bench-psud --simulate. Out-of-tree drivers load by module:Class. See docs/adding-a-driver.md.

Standard-library Python ≥ 3.11 only: no pyserial, no hidapi, nothing to install on the host but python3.

Prior art

Several MCP servers for bench supplies existed when this was written (checked 2026-11-27), each for one vendor: labmcp-bench-psu (Rigol, Siglent and Aim-TTi), an MCP server for the OWON SPE6103, siglent-mcp (with gated output), and servers for the Rigol DP832 and the RS3005P. None of them combines one daemon that owns the port, setters verified by readback, host ceilings, a write token, temperature interlocks and a live display, which is what this project is for.

Related MCP server: devpsu-mcp

Disclaimer

Kiprim, OWON and PCsensor are trademarks of their respective owners. This project is an independent effort and is not affiliated with, endorsed by, or connected with Kiprim, Shenzhen Kepurui Business Co., OWON, Lilliput or PCsensor. The names are used only to identify the hardware this software works with.

  • No warranty. Provided AS IS under the Apache License 2.0 (see LICENSE); liability is disclaimed to the fullest extent permitted by law.

  • Tested on one supply. A Kiprim DC620S, firmware V2.3.0. Other firmware and other models in the family may answer differently; run benchctl probe first.

  • The TEMPer driver is written from the published protocol and has not yet been checked against a stick of our own; see docs/drivers/temper.md.

  • Assume write access is dangerous. Anyone or anything holding the token (including an AI agent connected over MCP) can switch the output on and change the setpoints within the host ceilings.

  • An interlock is not a battery protection system. It acts on a temperature probe that lags the cell, through a USB stick, a daemon and a supply that can each fail. It is one more layer, not the only one.

Safety rules

These hold for the software and for the people using it.

  1. The daemon never switches the output on by itself. It switches the output off only on a clean stop with output_off_on_exit, or when a configured interlock trips. Starting, stopping, restarting or reconnecting sends queries only; a crash leaves the supply as it was. output_off_on_exit (config, default off) switches the output off when the daemon stops cleanly. An unexpected error in the worker is logged and the port reopened; if a thread dies anyway the process exits non-zero and systemd restarts it (queries only again).

  2. Identity first. Until the driver has confirmed the device's identity (owon-scpi: *IDN? matches expect_idn_prefix, default KIPRIM,DC620S, and, if set, expect_serial), nothing else is sent to the port.

  3. Host ceilings (max_voltage, max_current, default 15 V / 10 A) are checked before anything is sent, and must fit inside the hardware range the driver declares. The output is not switched on while a setpoint (perhaps set from the front panel) is above a ceiling.

  4. Protection limits keep a margin. The voltage setpoint must stay at least ovp_margin (0.1 V) below OVP and the current setpoint at least ocp_margin (0.1 A) below OCP. A request that would break that is refused, not reordered. Switching the output on is refused too while front-panel settings break it.

  5. Current and limit changes cycle the output. With the output on, a change to the current setpoint, OVP or OCP goes output OFF → write → output ON, for drivers that ask for it (owon-scpi does). Why: the two drivers that run this firmware family on a bench (lab_Kiprim, kibarra-esp32) report that OCP and the CURR:LIM value are one register and that writing it with the output on and current flowing near the new value tripped OCP. On a DC620S (V2.3.0) CURR and CURR:LIM read back separately and a live CURR write with open terminals left CURR:LIM alone, but nothing has been tested with current flowing, so the cycle stays the default. --live / "live": true skips it on explicit request. If a write fails mid-cycle, the output is left off. Voltage setpoint changes apply live.

  6. One process owns the port (flock + TIOCEXCL on the tty, flock on the hidraw node); the CLI's --direct mode fails while the daemon runs.

  7. Writes need a bearer token, generated on the host by the installer and readable only by root and the service. Without a token file, writes are disabled. Reads are open.

  8. Never connect a battery to the supply directly. The DC620S manual says the supply is not designed for direct battery charging: it has no reverse-current blocking and no remote sense, and its protections act on the supply, not the cells. Battery work needs a blocking element, a fuse, an independent meter and a dead-man disconnect, none of which this project provides.

  9. Never raise OVP to silence an OVP trip. *RST is never sent by this software.

  10. The probe cleans up. If a probe fails part-way, it still sends SYST:LOCK OFF, SYST:LOC and OUTP OFF for whatever it had turned on.

  11. Test setters, and interlocks, with open terminals or a resistive load, never a battery.

  12. Interlocks latch. A tripped interlock holds the output off and refuses output-on; after the condition clears it stays latched until a person resets it, and a reset is refused while the condition still holds. A reset switches nothing on. With interlocks configured, --direct refuses output-on (only the daemon reads the sensors).

Interlocks

An interlock watches one named temperature channel:

[[interlock]]
name = "pack-temperature"
channel = "pack"        # a channel name from a [[sensor]] table
max_c = 45.0            # trip above
min_c = 0.0             # trip below
stale_after = 5.0       # no fresh reading for this long counts as a trip
  • Trip when the channel reads above max_c, below min_c, or has had no fresh reading for stale_after seconds (sensor unplugged, driver error). If the output is on, the daemon sends output OFF at once through the driver (verified by readback, retried; a failure is logged as interlock_output_off_failed with "level": "critical"). While the condition holds, an output switched on at the front panel is switched off again.

  • Latched: output-on is refused (HTTP 409 "kind": "interlock", an MCP error, a CLI error) while the condition holds and afterwards, until someone resets it: benchctl interlock reset, POST /api/interlocks/reset or the MCP tool reset_interlocks. A reset is refused while the condition still holds.

  • At daemon start an interlock is waiting: output-on is refused until its channel has a first reading, but nothing is switched off before stale_after has passed without one, so a restart with a load running does not drop it because the sensor is slow to answer.

  • An interlock on a channel no [[sensor]] provides is a config error: the daemon will not start.

  • The supply's own OVP/OCP/OTP and the host ceilings are unchanged and independent of this.

States in the status: waiting, ok, tripped, latched. Journal events: interlock_armed, interlock_trip, interlock_clear, interlock_reset, interlock_output_off, interlock_output_off_failed.

Install (Debian 12/13 host)

Needs python3 ≥ 3.11 and nothing else. From a checkout on the host, as root:

install/install.sh install            # detects the CH340 tty, writes the udev rule
install/install.sh install --tty /dev/ttyUSB0      # if more than one CH340 is present
install/install.sh install --allow-from 192.0.2.0/24  # also open the port in nftables (see below)
install/install.sh install --display  # also the live display on the host's monitor (tty8)
install/install.sh upgrade            # same as install; keeps config, token, udev port and display
install/install.sh upgrade --no-restart  # install, but leave a running bench-psud on its old code
install/install.sh uninstall [--purge]

Or from a workstation, deploying the committed HEAD over SSH:

scripts/deploy.sh <ssh-host> install

What it installs:

Path

What

/usr/local/lib/bench-psu/ai_bench_psu/

the package

/usr/local/bin/bench-psud, benchctl, benchtop

launchers (python3 -I)

/usr/local/bin/kiprim

the CLI under its earlier name, for scripts; one notice line on stderr

/usr/local/lib/bench-psu-drivers/

searched for out-of-tree drivers; never written by the installer

/etc/bench-psu/config.toml

config, root:bench-psu 0640, installed once, never overwritten

/etc/bench-psu/token

write token, root:bench-psu 0640, generated once, never printed

/etc/udev/rules.d/70-bench-psu.rules

/dev/kiprim from the supply's USB port path (ID_PATH), ID_MM_DEVICE_IGNORE=1; the TEMPer data interfaces to group bench-psu, mode 0660, /dev/temperN

/etc/sysusers.d/bench-psu.conf

system user bench-psu, member of dialout

/etc/nftables.d/bench-psu.nft

only with --allow-from

/etc/systemd/system/bench-psud.service

the unit, enabled, sandboxed (ProtectSystem=strict, DevicePolicy=closed + char-ttyUSB + char-hidraw)

/etc/systemd/system/benchtop.service

only with --display: the live display on a virtual console, enabled

CH340 bridges carry no USB serial number, so the supply is identified by the port it is plugged into: move it to another USB port and re-run install.sh install --tty /dev/ttyUSBn.

On Debian, remove brltty if it is installed (it grabs CH340 devices).

--allow-from <CIDR> writes /etc/nftables.d/bench-psu.nft (add rule inet filter input … tcp dport <port> accept) and restarts nftables after a syntax check. It assumes an nftables.conf with an inet filter table, an input chain and include "/etc/nftables.d/*.nft"; on other firewalls, open the port yourself.

To try the hardware before installing anything, scripts/bundle.py prints a single-file copy of the CLI that runs from memory: python3 scripts/bundle.py | ssh <host> python3 - --direct --device /dev/ttyUSB0 probe.

Upgrading from bench-psu 0.2 / 0.3

This project continues bench-psu under a device-neutral name. install.sh upgrade on a bench-psu host keeps the unit name (bench-psud), the user, /etc/bench-psu (config and token) and the supply's USB port: it reads the ID_PATH from 70-kiprim.rules into 70-bench-psu.rules, replaces kiprim-top.service with benchtop.service on the same console, and installs kiprim as a compatibility launcher. A config with [serial] is read as [psu] with driver = "owon-scpi". /api/status keeps every key it had and adds temperatures, sensors and interlocks.

Live display

benchtop (also benchctl top) is a full-screen view for a screen at the bench, in the style of nvtop: big volts (yellow) and amps (cyan); a chart of both over the last 10 minutes with volts on the left axis and amps on the right (like the supply's own front screen); a band with output ON/OFF, CV/CC, the output-on time, power, the OVP/OCP/OTP trip flags and one chip per temperature channel; then setpoints and protection limits, the Wh / Ah counters and the daemon's link health.

  • The V/A chart draws the voltage setpoint and the current limit as dotted lines while the output is on, marks each setpoint change with a ▼ on its top border, and has a mode strip under it (CV green, CC amber, output off grey, fault red).

  • Fine scale: when the measured values on an axis move less than a quarter of its full scale (0 to a round top), that axis spans just the data on round steps, at least 50 mV or 5 mA wide, so a few millivolts or milliamps of drift are visible. The legend says ↕ fine scale V+A, the axis labels gain a digit, and a setpoint line outside the fine range is not drawn.

  • Under each big readout, while the output is on, a trend: the least-squares slope in mV/min and mA/min over the last 10 minutes, restarting when the output goes on or a setpoint changes (it needs 30 s of data first) and cleared when the output goes off.

  • With sensors configured, a temperature chart under the V/A chart on the same time axis, one line per channel, with each interlock's max_c and min_c as dashed red lines. Channel chips turn amber within 5 °C of an interlock's max_c and red when the reading is not fresh or the interlock is tripped or latched. A red INTERLOCK banner says what tripped and how to reset it.

  • + and - change the chart span (1 minute to 6 hours); q quits.

It reads GET /api/status and nothing else: no token, no writes, no device, so it can run anywhere (BENCH_URL=http://<host>:8100 benchtop). The charts are the display's own memory of what it has read, so they start empty when the display starts. When the daemon cannot be reached, a red NO LINK banner says so and the last reading turns red with its time. A red STALE or NO LINK banner also appears when the daemon itself has had no fresh poll from the supply for --stale-after seconds (default 3). benchtop --once prints one frame as text and exits 1 when the reading is not fresh.

install.sh install --display runs it at boot on a local monitor: benchtop.service on virtual console 8 (--display-vt N for another), outside the login consoles tty1–6, which keep their prompts (Ctrl+Alt+F1 to log in, Alt+F8 back to the display). It sets the Lat15-TerminusBold32x16 console font if the distribution has it (120×37 on a 1920×1200 monitor), runs as a dynamic user that may reach only localhost, and systemd restarts it if it stops. --no-display removes it.

Config

/etc/bench-psu/config.toml, see install/config.toml. Defaults:

Key

Default

Meaning

[psu] driver

owon-scpi

the PSU driver; its own options follow in the same section

[psu] device (owon-scpi)

/dev/kiprim

stable symlink from the udev rule

[psu] expect_idn_prefix / expect_serial (owon-scpi)

KIPRIM,DC620S / ""

identity required before anything else is sent

[psu] reply_timeout / setter_error_window / lost_after (owon-scpi)

0.5 / 0.15 / 3

line timing

[psu] reconnect_interval

2.0

seconds between reopen attempts

[[sensor]] name, driver, interval, channels

— / — / 1.0 / the driver's ids

one table per sensor device; channels maps driver channel ids to names; other keys are the driver's options

[[sensor]] stale_after

max(3, 3 × interval)

a reading older than this is not ok (its celsius is null)

[[interlock]] name, channel, max_c, min_c, stale_after

— / — / none / none / 5.0

see Interlocks

[poll] interval

0.25

4 Hz measurement poll, fixed rate

[poll] stale_after

3.0

no successful poll for this long: link stale, /api/health 503

[poll] sample_interval_on / _off

2.0 / 60.0

sample journal lines while the output is on / off (0 = never)

[poll] setpoint_interval

5.0

full refresh of setpoints and limits

[http] listen / port

0.0.0.0 / 8100

API, status page, MCP

[http] write_allow_from

[]

optional CIDR allow-list for writes, on top of the token

[http] allowed_hosts

[]

extra Host names accepted (IP literals, localhost, the host's own name and FQDN always are)

[http] request_timeout

10.0

an idle or stalled connection is dropped after this

[http] mcp

true

MCP endpoint at /mcp

[limits] max_voltage / max_current

15.0 / 10.0

host ceilings for setpoints

[limits] ovp_margin / ocp_margin

0.1 / 0.1

V / A the protection limits must stay above the setpoints

[behaviour] output_off_on_exit

false

switch the output off when the daemon stops

The plain HTTP API carries the token in the clear on the LAN. Put a TLS proxy in front, or use write_allow_from, if that matters on your network. Every request, read or write, must carry a Host header that is an IP literal or one of the host's names, and a browser Origin must match it: that stops a web page from reaching the daemon through DNS rebinding.

HTTP API

Reads are open. Writes are POST with a JSON body and Authorization: Bearer <token>. Every write response carries ok, result (the read-back values) and the fresh status.

Method and path

Body

Does

GET /

status page (refreshes every 2 s)

GET /api/status

the state document below

GET /api/health

200 while the supply's link is up, 503 otherwise

POST /api/set

{"volts": 13.8, "amps": 4, "live": false}

both, at most one output cycle

POST /api/voltage

{"volts": 13.8}

voltage setpoint (live)

POST /api/current

{"amps": 4, "live": false}

current limit (cycles the output if on)

POST /api/output

{"on": true}

output on/off

POST /api/ovp

{"volts": 15, "live": false}

over-voltage protection limit (cycles the output if on)

POST /api/ocp

{"amps": 5, "live": false}

over-current protection limit (cycles the output if on)

POST /api/lock

{"locked": true}

front-panel lock

POST /api/probe

{"level": "queries"}

the driver's conformance probe

POST /api/counters/reset

{}

reset the Wh / Ah counters (sends nothing to the supply)

POST /api/interlocks/reset

{}

reset latched interlocks (refused while a condition holds; switches nothing on)

Errors: 401/403 auth, 409 conflict with present setpoints, an interlock ("kind": "interlock") or a refused probe, 422 above a ceiling or the hardware range, 501 the supply has no such feature, 502 the supply answered ERR or read back a different value, 503 supply not connected or the wrong device, 504 timeout. On 401/403/422 nothing is sent to the supply; on 409 the daemon has read the present values (queries) but sent no setter.

/api/status:

{
  "time": "2026-11-11T20:00:00.000+00:00",
  "identity": {"manufacturer": "KIPRIM", "model": "DC620S", "serial": "…", "firmware": "V2.3.0", "raw": "…"},
  "output": true, "mode": "cv", "mode_code": 1,
  "setpoints": {"voltage": 13.8, "current": 4.0},
  "protection": {"ovp_limit": 63.0, "ocp_limit": 21.0, "ovp_tripped": false, "ocp_tripped": false, "otp_tripped": false},
  "measured": {"voltage": 13.8, "current": 0.0, "power": 0.0},
  "energy": {"wh": 0.0, "ah": 0.0, "output_on_s": 0.0, "since": "…"},
  "front_panel_lock": null,
  "link": {"state": "ok", "ok": true, "device": "/dev/kiprim", "age_s": 0.1, "poll_hz": 4.0,
           "reconnects": 0, "worker_alive": true, "…": "…"},
  "ceilings": {"max_voltage": 15.0, "max_current": 10.0, "ovp_margin": 0.1, "ocp_margin": 0.1},
  "warnings": [],
  "daemon": {"version": "0.4.0", "started": "…", "output_off_on_exit": false, "writes_enabled": true,
             "driver": "owon-scpi"},
  "temperatures": {"pack": {"celsius": 24.3, "last_celsius": 24.3, "sensor": "temper", "channel": "probe",
                            "age_s": 0.4, "ok": true},
                   "air":  {"celsius": 21.0, "last_celsius": 21.0, "sensor": "temper", "channel": "internal",
                            "age_s": 0.4, "ok": true}},
  "sensors": {"temper": {"driver": "temper", "identity": {"firmware": "…", "…": "…"}, "interval": 1.0,
                         "channels": {"internal": "air", "probe": "pack"},
                         "link": {"state": "ok", "ok": true, "age_s": 0.4, "last_error": null, "connects": 1}}},
  "interlocks": [{"name": "pack-temperature", "state": "ok", "ok": true, "channel": "pack", "max_c": 45.0,
                  "min_c": 0.0, "stale_after": 5.0, "since": null, "reason": null}]
}

mode is standby, cv, cc or fault. A temperature's celsius is null unless the reading is fresh (ok); last_celsius is the last value read, however old. The document is shaped for a Home Assistant RESTful sensor (value_template: "{{ value_json.measured.voltage }}" and so on). tests/test_status_compat.py checks that every key of a bench-psu 0.2.0 status document is still there.

MCP

POST /mcp is a Model Context Protocol server (Streamable HTTP transport, JSON responses, no SSE stream), built into the daemon so the one-owner rule still holds. Without the token it offers only get_status; with Authorization: Bearer <token> it adds set_voltage, set_current, set_voltage_current, set_output, set_ovp, set_ocp, set_lock, reset_counters, reset_interlocks and probe, with the same ceilings, rules and interlocks as the HTTP API. A wrong token is refused (401/403), not quietly treated as read-only. Its server instructions repeat the safety rules for the agent, including never resetting an interlock unless the human asked after checking the cause. Tested with the official MCP Python SDK client.

claude mcp add --transport http bench-psu http://<host>:8100/mcp     # read-only

For read-write, add --header "Authorization: Bearer <token>", taking the token from /etc/bench-psu/token on the host without echoing it into shared logs.

CLI

benchctl status                 full state, with temperatures and interlocks
benchctl idn                    identity string
benchctl set 13.8 4 [--live]    voltage and current limit
benchctl volt 12.5              voltage only
benchctl curr 2 [--live]        current limit only
benchctl on | off               output
benchctl ovp 15 | ocp 5 [--live]  protection limits
benchctl reset-counters         zero the daemon's Wh / Ah counters
benchctl lock | unlock          front panel
benchctl temps                  temperatures, sensors, interlocks
benchctl interlock [reset]      interlock states, or reset latched ones
benchctl watch [-i 1]           measurements until Ctrl-C
benchctl top [--window 10m]     the live display (same as benchtop); read-only
benchctl probe [--level L] [--markdown]
benchctl --direct status        bypass the daemon (it must be stopped)
benchctl --json status          JSON output

On the host as root the CLI reads the token file itself, so ssh <host> benchctl set 13.8 4 works. Elsewhere set BENCH_URL=http://<host>:8100 and BENCH_TOKEN (KIPRIM_URL and KIPRIM_TOKEN still work). kiprim … is benchctl … with a one-line notice on stderr: same subcommands, same stdout, same exit codes. It goes away in a later release.

Probe levels (owon-scpi)

benchctl probe runs the driver's protocol table and prints a conformance table. Each level includes the previous one.

Level

Sends

Needs

queries

every query; changes nothing

nothing

setters

each setpoint and limit written back with its present value, SYST:LOCK ON/OFF (read back with SYST:LOCK? if it exists), SYST:REM/SYST:LOC

output off

output

OUTP ON, mode read, OUTP OFF

output off; open terminals or a resistive load

ocp_quirk

OUTP ON, live CURR and CURR:LIM writes of the unchanged values, checking for an OCP trip after each, OUTP OFF; only meaningful with current flowing

as above

The command table and what the DC620S answered are in docs/drivers/owon-scpi.md.

Development

uv sync
uv run pytest
uv run bench-psud --simulate --listen 127.0.0.1 --config install/config.toml   # simulated supply

CI (GitHub Actions) runs the tests on Python 3.11–3.13 and shellcheck on the shell scripts.

The simulator (ai_bench_psu.drivers.psu.owon_scpi.simulator) models the verified behaviour, with switches for the unverified quirks (ocp_quirk, three-field MEAS:ALL?, digit OUTP?, missing MEAS:ALL:INFO?, SYST:LOCK?, dropped replies, readback offsets) and a resistive load. sim-temp is a thermometer whose values, absence and failures the tests set.

References

No code was copied from these projects; they informed the command tables, the report layouts and the quirk list.

License

This project is licensed under the Apache License, Version 2.0.

Copyright 2026 TechBlueprints

Licensed under the Apache License, Version 2.0 (the "License");
you may not use this file except in compliance with the License.
You may obtain a copy of the License at

    http://www.apache.org/licenses/LICENSE-2.0

Unless required by applicable law or agreed to in writing, software
distributed under the License is distributed on an "AS IS" BASIS,
WITHOUT WARRANTIES OR CONDITIONS OF ANY KIND, either express or implied.
See the License for the specific language governing permissions and
limitations under the License.

Tool Schema Changelog

Recent tool additions, removals, and schema changes observed during successful MCP inspections.

Loading tool schema changelog…

    Related MCP Connectors

    Related MCP Servers

    • F
      license
      Not graded
      quality
      D
      maintenance
      MCP server for programmable bench power supply control, enabling autonomous bootloader-entry probing via precise power cycling and telemetry.
      1
      -
    • A
      license
      A
      quality
      D
      maintenance
      MCP server for controlling B&K Precision and Manson bench power supplies, enabling voltage/current setting, output switching, and power cycling.
      8
      3
      MIT
    • A
      license
      Not graded
      quality
      D
      maintenance
      MCP server for reading and controlling an OWON SPE6103 programmable DC power supply over USB serial, with safety checks and verified SCPI commands.
      MIT
    • F
      license
      B
      quality
      B
      maintenance
      Enables MCP clients to list PD profiles, probe USB-C devices, audit PDOs, read live power telemetry, trigger PD/PPS voltages with safety defaults, and reset to safe 5V. It runs locally over stdio JSON-RPC with strict tool schemas and dry-run defaults for mutating operations.
      6
      -