openedu09

Open-source driver, GUI and MCP server for the Whadda WSEDU09 (formerly Velleman EDU09) USB PC oscilloscope kit. Runs on Linux and Windows (Python 3.12+), with a simulator for hardware-free use.

CI License: MIT Python 3.12+

openedu09 GUI showing a simulated 1 kHz square wave

The GUI running against the built-in simulator (--simulate).

openedu09 spectrum analyzer showing a 30 kHz 60 % duty square wave at 20 us/div

View > Spectrum (Ctrl+F): flat-top window, 8-frame average, peak labels, THD and the ETS note (simulated 30 kHz square wave, 60 % duty, ETS timebase).

Naming: the project is called openedu09, the Python import package is edu09, and the commands are edu09-gui, edu09-mcp, edu09-calibrate, edu09-gain and edu09-probe.

Features

  • Driver library (edu09.Scope): configure, single shot, continuous run, volts calibrated per unit; 15 timebases from 500 ms/div to 10 us/div (50/20/10 us use equivalent-time sampling); AUTO / NORMAL / off trigger with rising or falling edge; six ranges, 0.1 to 5 V/div.

  • GUI (edu09-gui, PySide6 + pyqtgraph): live trace, cursors, measurements, CSV export, auto-set, spectrum analyzer, calibration wizard, optional built-in MCP server.

  • MCP server (edu09-mcp): lets an AI agent configure the scope, acquire, measure, view the spectrum and export waveforms, over stdio or streamable HTTP.

  • Calibration wizard (GUI, edu09-calibrate, MCP calibration_* tools): calibrate your own unit with a multimeter and the DC sources you have (batteries, USB 5 V, a bench supply): zero, optional RV1 trimmer adjustment, per-range gain on both sides of ground, verify.

  • Spectrum analyzer (GUI, MCP spectrum, edu09.analysis.spectrum): 1024-sample FFT in Vrms/dBV with five windows, peaks, THD, averaging, and aliasing/ETS/clipping warnings.

  • Simulator (--simulate): everything works without hardware.

  • Safety: one process owns the device at a time (file lock); calibration is opt-in and needs the input disconnected; edu09-probe blocks unknown opcodes.

Related MCP server: keyscope-mcp

Quick start

git clone https://github.com/BonBear/openedu09.git && cd openedu09
python3 -m venv .venv && . .venv/bin/activate
pip install -e '.[gui]'
  • Windows: use py -3.12 -m venv .venv and .venv\Scripts\activate (Python 3.12 or 3.13 only, no driver needed).

  • Linux: install the udev rule once, and libxcb-cursor0 for the GUI on X11:

    sh packaging/linux/install-udev.sh        # uses sudo; then replug the scope
    sudo apt install libxcb-cursor0           # Debian/Ubuntu, others: docs/install.md

Details and troubleshooting: docs/install.md. Then:

edu09-gui                               # GUI on the attached scope
edu09-gui --simulate "sine:1khz:2vpp"   # GUI without hardware
edu09-mcp                               # MCP server on stdio
edu09-mcp --simulate                    # MCP server on a simulated scope
edu09-mcp --http                        # MCP server on http://127.0.0.1:8765/mcp

MCP with Claude Code

Add the server to .mcp.json (use the full path to the venv's executable; on Windows ...\.venv\Scripts\edu09-mcp.exe), or run claude mcp add edu09 -- /path/to/openedu09/.venv/bin/edu09-mcp:

{
  "mcpServers": {
    "edu09": { "command": "/path/to/openedu09/.venv/bin/edu09-mcp" },
    "edu09-sim": {
      "command": "/path/to/openedu09/.venv/bin/edu09-mcp",
      "args": ["--simulate", "square:1khz:3vpp"]
    }
  }
}

Only one process can own the scope. If the GUI is open (start it with edu09-gui --mcp-http, or Tools > MCP server), a plain stdio edu09-mcp detects it and forwards to the GUI automatically, so the same config works either way. You can also connect straight to the GUI's endpoint: claude mcp add --transport http edu09-gui http://127.0.0.1:8765/mcp.

Tools, options, the bridge, remote access (no authentication: trusted networks only) and debug recording are described in docs/usage.md.

Calibration

Run the wizard once per unit (GUI Tools > Calibrate, or edu09-calibrate); you need a multimeter and one or more DC sources.

  1. Turn the board trimmer RV1 to its middle position (first time only).

  2. Zero with the input disconnected.

  3. Optional trimmer step: a 1.5 V battery at 0.5 V/div; turn RV1 until the live reading matches the multimeter.

  4. Per-range gain steps with the voltages you have, then save (a backup is made) and verify.

edu09-calibrate --plan-only --fixed 1.5@10%,rev,label=AA --fixed 5,label=USB   # what you can cover
edu09-calibrate --fixed 1.5@10%,rev,label=AA --fixed 5,label=USB --trimmer

Details, sources, the 5 V/div range and the file format: docs/usage.md.

Safety

  • Max input 30 V (AC + DC), input impedance 100 kOhm (vendor manual). The driver cannot protect the hardware from over-voltage; the calibration wizard asks for at most 15 V by default.

  • The calibration's zero step needs the input disconnected. Only swap the leads of floating sources (batteries, isolated supplies), never of a source that shares ground with the PC.

  • Only one program can use the scope at a time; the second one gets a "device busy" message naming the owner. Do not run the vendor software at the same time.

Status and limitations

Alpha (0.1.0, released 2026-10-25). Firmware 1.01 of the WSEDU09/EDU09 is what was verified.

  • Verified on real hardware: Linux and Windows (Python 3.12) with a hardware test suite, the GUI, the MCP server and bridge, the calibration wizard (including the RV1 trim and the 5 V/div range with a 25 V source) and the spectrum analyzer (on a 555 square wave). The protocol is documented in docs/protocol.md.

  • Untested: macOS, other firmware versions, other Whadda/Velleman scopes. Single channel (CH1) only.

  • Vertical gain: an uncalibrated unit reads 0.4-2.6 % high, and large negative deflections (more than ~4 div below ground at 0.2/0.5 V/div) up to ~6 % too large. The wizard measures and corrects both per unit (verify within ~0.6 % on the development unit). Without a calibration file, typical offsets and no gain correction are used, and the scope is shown as "Uncalibrated" (GUI status bar, MCP get_status). The 5 V/div range needs a 20-30 V reference (--max-volts 25).

  • ETS (50/20/10 us/div) needs a repetitive, triggered signal; frames show a few isolated outlier samples (View > Despike).

  • A readout timeout at GUI startup was seen once and has not been reproduced; if it happens, run with --record FILE and report it.

Development

pip install -e '.[gui,dev]'
ruff check . && ruff format --check . && mypy && pytest -q

Hardware tests, safety rules and how to contribute captures: CONTRIBUTING.md. Security reports: SECURITY.md. Changes: CHANGELOG.md. Design documents: docs/specs/.

How it was built

The wire protocol was reverse engineered from the vendor software for the purpose of interoperability, then confirmed with captures from a real device. The code was written with a team of AI coding agents (see AGENTS.md) under human direction and review.

License

MIT, copyright (c) 2026 Björn Sjölén.

Disclaimer

This project is independent. It is not affiliated with or endorsed by Whadda or Velleman. "Whadda", "Velleman", "WSEDU09" and "EDU09" are trademarks of their respective owners. No vendor software, firmware, documentation or other vendor files are included in this repository. Protocol knowledge was obtained for interoperability purposes. Use at your own risk; the authors accept no liability for damage to hardware or equipment.

Related MCP Connectors

Related MCP Servers

  • A
    license
    A
    quality
    D
    maintenance
    An MCP server that enables AI assistants to control Siglent SDS oscilloscopes over a local network using SCPI commands. It allows users to measure signals, configure channel and acquisition settings, and capture waveforms or screenshots through natural language.
    13
    17 npm
    7
    MIT
  • A
    license
    Not graded
    quality
    D
    maintenance
    AI-native oscilloscope control. A pure-Python MCP service that lets LLM agents control Keysight EDUX1052G oscilloscopes via a compact DSL.
    MIT
  • A
    license
    A
    quality
    C
    maintenance
    An MCP server that enables AI agents to operate an oscilloscope through high-level tools like signal capture and measurement, abstracting vendor-specific SCPI commands.
    18
    4
    MIT
  • A
    license
    Not graded
    quality
    D
    maintenance
    An MCP server that connects AI assistants to the Owon DS1102 oscilloscope, enabling waveform capture, measurement retrieval, and remote control of scope settings.
    GPL 3.0