sample
Records timestamped multimeter readings at a fixed interval and returns statistics plus all readings, enabling timed measurement runs and post-run analysis.
Instructions
Record timestamped readings at a fixed interval and return statistics plus all samples.
Readings are paced by this computer. A reading takes about 0.3 s at speed slow and in most other functions, 0.6 s for frequency and period; with a shorter interval readings are taken back to back. Without function the present function and range are kept; otherwise they are set as in measure.
Args: duration_s: Recording time in seconds (max 600). interval_s: Time between readings in seconds (min 0.1). function: Measurement function. Leads in HI/LO for voltage, resistance_2w, capacitance, frequency, period, continuity, diode and temperature; in I/LO for currents; resistance_4w also uses HI Sense/LO Sense. range: Fixed range in V, A, Ω or F; omit for autorange. dc_voltage 0.6, 6, 60, 600, 1000; ac_voltage 0.6, 6, 60, 600, 750; dc_current 0.0006, 0.006, 0.06, 0.6, 6, 10; ac_current 0.06, 0.6, 6, 10; resistance_2w/4w 600, 6e3, 60e3, 600e3, 6e6, 60e6, 100e6; capacitance 2e-9, 20e-9, 200e-9, 2e-6, 20e-6, 200e-6, 0.01. Other functions have no range. speed: Only for dc_voltage, dc_current and resistance: slow (10 PLC, default, most accurate), medium (1 PLC) or fast (0.3 PLC). sensor: Required for temperature: pt100 (RTD) or thermocouple type B, E, J, K, N, R, S or T.
Input Schema
| Name | Required | Description | Default |
|---|---|---|---|
| range | No | ||
| speed | No | ||
| sensor | No | ||
| function | No | ||
| duration_s | No | ||
| interval_s | No |
Output Schema
| Name | Required | Description | Default |
|---|---|---|---|
No arguments | |||