A local stdio MCP server for OWON oscilloscopes. Connect an MCP client to read settings, capture waveforms, query measurements and the multimeter, export data, and optionally control acquisition or record a sequence of screen traces.

USB transport, protocol detection and waveform decoding come from OWON Scope Web, pinned to a source revision in pyproject.toml and uv.lock. This project adds MCP tools, explicit USB ownership, response limits, recording and an HDS200 control adapter.

Hardware validation: HDS2102S, firmware V4.7.1, on macOS / Apple Silicon. Other OWON models use upstream drivers and have not been validated here. See hardware support and troubleshooting for the tested controls and protocol caveats.

Install from source

Requires Python 3.11+, Git, uv, and libusb for real hardware. This repository is the installation source; the instructions do not depend on a PyPI release.

git clone https://github.com/darekxan/owon-mcp.git
cd owon-mcp
# macOS hardware access:
brew install libusb
uv sync --locked
uv run --no-sync owon-mcp --version

On Linux, install libusb-1.0 and configure permissions using the upstream USB setup instructions. On Windows, follow the upstream WinUSB driver instructions. The offline demo and test suite do not require a scope or libusb. Hardware compatibility on Linux and Windows is not established by the offline tests.

Related MCP server: ScopeMCP

Configure an MCP client

Use the installed executable as a local stdio server. Copy examples/mcp.json into a client that accepts mcpServers configuration, replacing both paths with your checkout location:

{
  "mcpServers": {
    "owon": {
      "command": "/absolute/path/to/owon-mcp/.venv/bin/owon-mcp",
      "args": ["--export-dir", "/absolute/path/to/owon-mcp/exports"]
    }
  }
}

On Windows, the executable is .venv/Scripts/owon-mcp.exe; use forward slashes or JSON-escaped backslashes in paths. An absolute executable path avoids dependence on a GUI client's shell environment. Restart the configured server after changes.

For Codex, examples/codex.toml provides the equivalent mcp_servers.owon entry. For a project-local configuration, copy it to .codex/config.toml and replace its paths. The local configuration is ignored by Git. The optional hardware control scripts also read this file.

The default profile disables instrument setting and acquisition writes. It still allows explicit USB discovery/connection, queries, and local capture exports. Add --allow-control to args to enable controls and recording. There are no unrestricted SCPI writes, USB-reset, firmware-flashing or function-generator tools.

The MCP server does not open a network port. Protocol output goes to stdout and logs to stderr. Starting it, initializing MCP, listing tools and reading status do not scan or open USB. There are no automatic connections or background scans.

First capture

  1. Release the scope in OWON Scope Web or any other USB application. Only one process can claim its interface at a time.

  2. Put the instrument in HID/control mode instead of MSC storage mode.

  3. Call owon_discover to list candidates without probing their protocol.

  4. Call owon_connect. Supply a discovered device_id if several scopes exist.

  5. Call owon_get_settings, then owon_capture with, for example:

    {"max_points": 300, "include_samples": true, "include_raw": false}
  6. Optionally call owon_export_capture with {"format": "json"} or {"format": "csv"}. Export before disconnecting: disconnect clears the cached capture.

  7. Call owon_disconnect to release USB. Normal server shutdown also disconnects.

You can ask a client: “Discover my OWON scope, connect, read its settings, capture both enabled channels, export the full capture as JSON, then disconnect.” Discovery and connection happen only when those tools are explicitly called.

Tools

Tool

Purpose

owon_status

Cached connection, demo, control and recording state

owon_discover

List USB candidates without protocol probing

owon_connect / owon_disconnect

Claim and identify / release one instrument

owon_capabilities

Cached driver capabilities and accepted control values

owon_get_settings

Read channels, timebase, trigger and acquisition state

owon_capture

Capture enabled traces and statistics, with optional samples/raw data

owon_dmm_read

Query multimeter values without changing its mode

owon_scope_measurements

Query instrument-calculated extrema, amplitude, period or frequency

owon_control

Typed setting/acquisition writes; requires --allow-control

owon_record_start

Start bounded background recording; requires --allow-control

owon_record_status / owon_record_stop

Check progress / request stop and cleanup

owon_get_raw

Read cached upstream diagnostics

owon_export_capture

Write a full cached capture as JSON or CSV

Resources owon://status and owon://capture/latest expose cached status and a capture summary. See tool reference for arguments, limits, control values, cache lifetime and data semantics.

Recording

After explicitly connecting, set the channels, probe ratios and timebase, then call owon_record_start:

{"duration_s": 60, "interval_s": 0.2, "filename": "sequence.json"}

It selects AUTO sweep and RUN, saves full traces and raw diagnostics under --export-dir, and returns immediately. Poll owon_record_status or call owon_record_stop. Recordings are limited to 600 seconds, intervals of 0.05–5 seconds, and 10,000 scheduled frames. Slow USB reads can reduce the achieved rate.

Recording changes the instrument state: cleanup attempts STOP and disconnect on completion, manual stop or failure. It does not restore the previous sweep or run state. Reconnect before further scope work. A stopping result means cleanup is still pending; inspect the terminal status and any error/cleanup_errors. Only status and recording stop are available while recording is active.

Limits that matter when interpreting data

  • A capture is a sequence of header/channel reads during acquisition, not an atomic hardware snapshot. Recordings have gaps and do not extend ADC memory.

  • Voltage scaling, probe metadata and point spacing come from the pinned upstream decoder. Screen-point spacing is not necessarily the acquisition sample period.

  • max_points decimates each returned trace; it is not a peak-preserving filter. Full samples stay cached for export. include_raw=true attaches full diagnostics, which are outside the sample limit and can produce large responses.

  • Bare numeric DMM replies have unit=null; units and overload text are never inferred. Activate DMM mode on the instrument when needed.

  • Most controls confirm command dispatch only. HDS200 sweep/Single readback is checked; use settings readback to assess other controls. Native position offsets are not advertised as calibrated divisions.

  • Automatic USB reset recovery is disabled. A timeout is reported to the client; manual reconnection may be necessary.

Try it without hardware

Add --demo to the MCP client's server arguments to use simulated traces, settings, DMM values and measurements. Discovery and connection remain explicit, and status and captures identify the simulation. Add --allow-control for simulated writes. Running the executable directly starts stdio MCP and waits for a client:

uv run --no-sync owon-mcp --demo

Development and checks

uv sync --locked
uv run --no-sync pytest -q
uv run --no-sync ruff check .
uv build

Tests use synthetic drivers, replay data and real MCP clients, including a stdio subprocess. They do not access hardware. GitHub Actions runs the offline checks and builds across the configured Python/platform matrix.

For an explicit hardware read check, disconnect other USB clients, then run:

uv run --no-sync python scripts/live_smoke.py

It connects, reads settings/capture, exports and disconnects, without scope control writes. Add --dmm only when DMM mode is ready. Control and trigger checks are documented separately in hardware support; they change settings.

Local recordings, captures, firmware and client configuration are excluded from the repository and package archives.

CLI options

Option

Default / meaning

--demo

Use a USB-free simulated instrument

--allow-control

Enable setting/acquisition writes and recording

--export-dir PATH

Output directory; defaults to ./exports in the process working directory

--verbose

Send diagnostic logs to stderr

--help / --version

Print usage/version without loading the USB backend

License and upstream

MIT; see LICENSE. OWON Scope Web is MIT-licensed by Francisco Mesas Cervilla; see NOTICE. The upstream revision is cb86012000048b8441318a293fe7a41a83f98023. This project is independently developed and is not affiliated with OWON. Firmware is not included.

Tool Schema Changelog

Recent tool additions, removals, and schema changes observed during successful MCP inspections.

Loading tool schema changelog…

    Related MCP Connectors

    Related MCP Servers

    • A
      license
      Not graded
      quality
      D
      maintenance
      MCP server for controlling Hantek DSO2D15 and other DSO2000-family oscilloscopes via USB and SCPI, enabling waveform acquisition, measurements, and screen captures.
      1
      MIT
    • A
      license
      Not graded
      quality
      D
      maintenance
      An MCP server that connects AI assistants to the Owon DS1102 oscilloscope, enabling waveform capture, measurement retrieval, and remote control of scope settings.
      GPL 3.0
    • A
      license
      A
      quality
      B
      maintenance
      Enables MCP clients to discover tools, configure a Rigol oscilloscope over LAN, take measurements, and retrieve waveform captures via a containerized stdio server.
      21
      MIT
    • A
      license
      A
      quality
      C
      maintenance
      Enables MCP-compatible AI clients to operate a RIGOL MHO98 oscilloscope over USB, including configuration, waveform and measurement acquisition, FFT analysis, and protocol decoding.
      16
      MIT