# scpi-mcp

**LabLink gives agents a SCPI wire; scpi-mcp gives agents oscilloscope expertise.**

`scpi-mcp` is a [Model Context Protocol](https://modelcontextprotocol.io) server
that lets an agent *operate an oscilloscope* โ€” not just push SCPI strings at one.
It exposes goal-level tools (capture this signal, measure that, characterize a
channel) backed by a clean, instrument-agnostic interface. All vendor SCPI lives
in the underlying instrument library; the MCP layer stays a thin, uniform wrapper
with **no raw-SCPI escape hatch**.

```
        .-"""-. 
       / .===. \      Skippi  ๐Ÿงšโ€โ™‚๏ธ
       \/ 6 6 \/      the artistic scope gremlin
       ( \___/ )      "every waveform is a portrait of the magic pixies"
    ___ooo___ooo___
```

> **Skippi** is the mascot โ€” the artistic gremlin who lives in the scope and
> paints pictures of the magic pixies (the ones who carry every electron down
> the probe). When you see a clean trace, that's Skippi's latest masterpiece.

## Scope

First instrument: **Rigol DS1054Z** (unlocked to DS1104Z base) โ€” 4 analog
channels, 100 MHz. No `:SOURce` (function generator) or `:LA` (logic analyzer)
support. Reachable over USB or LAN (`TCPIP::<ip>::INSTR`).

## Architecture

- **`transport/`** โ€” owns discovery. Enumerates USB, falls back to best-effort
  LAN auto-discovery, then to an IP prompt; resolves a single VISA *resource
  string* and hands it to the instrument layer. Discovery never leaks upward.
- **`instruments/`** โ€” vendor-specific backends behind one abstract `base.py`
  interface. `rigol_ds1000z.py` is a thin wrapper over the (complete) library;
  `mock.py` implements the same interface with zero hardware. Adding an
  instrument is one new file here and **zero** changes to `tools/`.
- **`tools/`** โ€” instrument-agnostic MCP tools that call only `base.py`.
- **`config.py`** โ€” permission tiers (`read_only` / `read_config` / `full`),
  enforced at the tool layer. The *server* refuses; the model never decides.

## Status

**Live hardware, active bench use.** The Rigol DS1054Z is connected and
scpi-mcp is driving real measurements โ€” see [Skippi in the Wild](#skippi-in-the-wild)
for captures. The mock backend remains available for development and CI.

## Quickstart

```bash
uv sync
uv run pytest          # all green against the mock, zero hardware
uv run scpi-mcp        # start the MCP server (mock backend by default)
```

## Skippi in the Wild

Real captures from the bench, analyzed live via scpi-mcp.

### Power-off โ€” relay coil with freewheel diode

**The setup:** A 24 V relay coil (with freewheel diode installed) powered from a
bench supply. Probed at the moment of power-off to observe inductive kickback.
CH1 watches the supply rail; CH2 sits across the freewheel diode.

**Prompts used:**

```
Grab the scope screen.
What are the min and max voltages on each channel?
Is that spike on CH2 actually 60 V, or is the scale messing with me?
```

scpi-mcp called `capture_screen` and `measure_snapshot`, then explained the
visual illusion: CH2 is set to 1 V/div, so its 5.36 V kickback spike fills
5+ of the 8 vertical divisions and *looks* enormous. CH1 is at 8 V/div, so
the 24 V rail only occupies ~3 divisions despite being far larger in absolute
terms. No 60 V spike โ€” the freewheel diode clamped it cleanly to
V_supply + V_f โ‰ˆ 24.7 V, right where it should be.

![Power-off capture โ€” relay coil, freewheel diode installed](docs/images/scope_poweroff_long.png)

*500 ms/div (5 s full sweep). CH1 yellow, 8 V/div: 24 V supply decaying from
26.9 V โ†’ 2.24 V. CH2 cyan, 1 V/div: freewheel diode kickback, max 5.36 V,
min โˆ’2.08 V.*

---

### Power-off โ€” bare coil, no freewheel diode

**The setup:** The same 24 V relay coil, but with the freewheel diode *removed* โ€”
a bare inductor. Powered from a bench supply (100 nF across the rail to tame
switch-mode ripple). Two 10x probes straddle the disconnect on the high side:
CH1 on the supply side, CH2 on the coil side. Power is cut by pulling the +
lead, and a single-shot trigger catches the unclamped kickback.

**Prompts used:**

```
Prime the scope for a single-shot on CH2 falling โ€” I'll pull the + lead.
We didn't clear the clipping โ€” reset V/div on CH2 and re-arm for another one-shot.
Got it โ€” grab the screencap and give me the true peak.
```

scpi-mcp armed a single-shot falling-edge trigger on CH2, watched the trigger
status, and grabbed the screen plus both channel waveforms the instant it fired.
This time the *scale fought back the other way*: the kick kept overflowing the
window โ€” it railed at ยฑ130 V, then again at โˆ’304 V โ€” so scpi-mcp stepped CH2
coarser to 100 V/div until the whole transient fit. With no freewheel diode to
absorb it, the collapsing field threw a **โˆ’436 V** spike with a **+240 V**
overshoot โ€” a ~676 Vpp ring, roughly **18x** the 24 V rail โ€” then bled off
exponentially back toward the rail over ~1 ms. Compare the clamped case above,
where the diode held the same coil to โ‰ˆ24.7 V.

![Power-off capture โ€” bare coil, freewheel diode removed](docs/images/scope_poweroff_bare_coil.png)

*200 ยตs/div. CH1 yellow, 20 V/div: supply side holds ~24 V with a small coupled
ring at the break. CH2 cyan, 100 V/div: bare-coil kickback โ€” peak โˆ’436 V,
+240 V overshoot (~676 Vpp), then exponential bleed-back. No freewheel diode;
contrast the โ‰ˆ24.7 V clamped result above.*

## License

MIT

TDQS

A3.9/5.0

Scored across 18 tools

Disambiguation5/5

Each tool targets a distinct oscilloscope operation. Acquisition, capture, measurement, and configuration tools have clear, non-overlapping purposes. Even similar tools like acq_run and acq_single are differentiated by continuous vs single-shot acquisition.

Naming Consistency4/5

Tools predominantly follow a consistent verb_noun pattern (e.g., set_channel, capture_screen). Minor deviations like 'identify' and 'self_config' slightly break the pattern but remain intuitive. Overall, naming is predictable and readable.

Tool Count5/5

18 tools comprehensively cover essential oscilloscope operations without redundancy. The count is well-scoped for a domain-specific server, providing enough granularity for complex workflows while remaining manageable.

Completeness4/5

Core lifecycle tasks (connect, configure, acquire, capture, measure) are fully covered. Advanced features like FFT or automated calibration are absent but not critical for basic SCPI control. Minor gaps exist but do not hinder primary use cases.