Python driver, command-line tool and MCP server for the B&K Precision 880 handheld LCR meter, over its mini-USB virtual COM port. Use it to script measurements, log readings to CSV, sort parts against a reference, or let Claude Desktop take and interpret readings directly.

  • Reads and changes every setting the remote interface offers (L/C/R/Z/DCR, D/Q/θ/ESR, 100 Hz to 100 kHz, 0.3/0.6/1 V, series/parallel, tolerance) and checks each change by reading it back, because the meter never acknowledges commands.

  • Works around firmware quirks found on real hardware (see Protocol notes).

  • Takes an exclusive lock on the port, so the CLI and the MCP server can share one meter without reading each other's replies.

  • Finds the meter by its USB adapter's serial number, so it is never confused with another CP210x adapter on the same machine.

Not affiliated with or endorsed by B&K Precision.

Requirements

  • A B&K Precision 880 (tested with firmware 1.11) and a mini-USB cable. The meter runs on its own battery or 12 V supply, not USB power, so it has to be switched on.

  • Python 3.11+ and uv (or pip).

  • macOS, where it is tested. The meter enumerates as a Silicon Labs CP2102 (/dev/cu.usbserial-*). The code uses only pyserial, so Linux (/dev/ttyUSB*) should work, but has not been tried.

Related MCP server: SCPI MCP

Install

git clone https://github.com/JC-000/bk880.git
cd bk880
uv sync --all-extras        # or: pip install -e '.[mcp]'
uv run bk880 idn            # B&K Precision,880,<serial>,1.11

The extra mcp dependency is only needed for the Claude Desktop server.

Finding the meter

With a single CP210x serial adapter attached, the port is auto-detected by sending it *IDN?. With several, auto-detect refuses rather than probe the others, and you name the meter's adapter instead. bk880 ports lists the adapters with their USB serial numbers and locations, without opening any:

bk880 ports            # /dev/cu.usbserial-123A45678  serial 123A45678  location 0-1.4
export BK880_SERIAL=123A45678   # your adapter's serial

Stock CP2102s all report USB serial 0001, and macOS names their ports in the order it finds them, so with two attached a replug can swap /dev/cu.usbserial-4 and /dev/cu.usbserial-0001. That rules out BK880_PORT (a device path). The meter's adapter can be reprogrammed to report the meter's own serial, so BK880_SERIAL finds it on any USB port. Without reprogramming, BK880_LOCATION (the hub port its cable is in, such as 0-1.4) works as long as the cable stays in that port. Order of precedence: --port, $BK880_PORT, $BK880_SERIAL, $BK880_LOCATION, auto-detect.

Variable

Selects the meter by

Example

BK880_PORT

device path

/dev/cu.usbserial-0001

BK880_SERIAL

USB serial of its adapter

123A45678

BK880_LOCATION

USB hub port

0-1.4

Giving the meter's adapter a unique USB serial

The CP2102's serial string sits in its configuration memory and can be rewritten over USB (Silicon Labs vendor request 0xFF, register 0x3704, the same request cp210x-program uses). With the meter as the only CP210x attached:

brew install libusb
uv run --with pyusb python scripts/cp210x_serial.py                    # read: chip, lock, serial
uv run --with pyusb python scripts/cp210x_serial.py --write 123A45678  # then replug

The script uses Homebrew's libusb path (/opt/homebrew/lib), which is Apple-silicon macOS; edit it elsewhere. It refuses unless exactly one CP210x is attached and it is an unlocked CP2102. The vendor and product IDs are not touched, so a bad write still leaves an adapter that enumerates and can be rewritten.

CLI

bk880 ports                                 # CP210x adapters, USB serials, locations (opens none)
bk880 idn                                   # B&K Precision,880,123A45678,1.11
bk880 config                                # current settings as JSON
bk880 set --primary C --freq 120 --level 1  # change settings (verified by read-back)
bk880 fetch -n 10                           # 10 readings as JSON lines + stats on stderr
bk880 log -o run.csv -i 0.5                 # CSV log every 0.5 s until Ctrl-C
bk880 tol on --range 5                      # capture the part in the socket as nominal, ±5 %
bk880 tol off --secondary D                 # leave tolerance mode, re-select D
bk880 record on | bk880 record | bk880 record off   # meter's own max/min/average
bk880 stream -o run.csv                     # passive log of the meter's auto-fetch output

stream and record are CLI-only on purpose: stream needs someone at the meter to press USB, and the meter's own recording statistics are unreliable (see the protocol notes), so the MCP server does not offer them.

Sorting parts against a reference

Put the reference part in the socket and run bk880 tol on --range 5. The meter takes that reading as the nominal. Then swap in each part to test: readings now have deviation_pct (percent off nominal) where the secondary used to be. bk880 tol off --secondary D turns it off again.

Passive logging (auto-fetch)

bk880 stream reads the settings, hands the meter back to the front panel and waits; give the meter's USB button one short press (holding it toggles the backlight) and it pushes a reading every cycle, about 2.2 per second, which stream writes to CSV without sending anything. You can keep using the front panel. Changing the function on the panel mid-stream leaves the CSV's parameter labels stale.

Streaming does not keep the meter awake: it still switches itself off after its auto power-off delay (5 minutes here). If readings stop for 5 seconds once they have started, stream prints a warning to stderr and keeps waiting, so turning the meter on and pressing USB again resumes the same log. For longer runs, set AoFF in the meter's utility menu to 15, 30 or 60 minutes or OFF, press a button now and then, or use bk880 log, which polls.

Values are SI base units: F, H, ohm (theta in degrees). A blank/null primary value means over range.

Python

from bk880 import BK880, resolve_port, summarize

with BK880(resolve_port()) as m:   # $BK880_PORT, $BK880_SERIAL, $BK880_LOCATION or auto-detect
    m.set_primary("L"); m.set_secondary("Q"); m.set_frequency(1000)
    print(summarize(list(m.read_many(20))))

Claude Desktop (MCP)

Add this to ~/Library/Application Support/Claude/claude_desktop_config.json and restart Claude Desktop (quit with Cmd-Q, then reopen; do the same after updating the code). Use absolute paths for uv (which uv) and the checkout, because Claude Desktop does not inherit your shell's PATH. The server looks the serial up on every call, so it follows a renamed port or a different USB port without a restart.

{
  "mcpServers": {
    "bk880": {
      "command": "/opt/homebrew/bin/uv",
      "args": ["run", "--directory", "/path/to/bk880", "bk880-mcp"],
      "env": { "BK880_SERIAL": "123A45678" }
    }
  }
}

Tools:

Tool

Does

bk880_identify

model, serial number, firmware, port

bk880_get_config

primary/secondary parameter, frequency, level, series/parallel, tolerance

bk880_configure

change any of the above

bk880_measure

N readings with mean/stdev/min/max (optionally each reading)

bk880_frequency_sweep

average readings at each test frequency, then restore the original

bk880_tolerance

capture the part in the socket as nominal, set the ±range; readings then carry deviation_pct

bk880_log_csv

log N readings to a CSV file and return the stats

The port is opened per call and the meter is returned to front-panel control (*GTL) afterwards, so the meter's buttons and the CLI still work between calls.

Protocol notes

From the 880 manual (pp. 45-54), plus behaviour observed on the hardware:

  • 9600 8N1, no flow control; send \n, replies end in \r\n.

  • Set commands and invalid commands return nothing. Errors only show on the LCD (E10 unknown command, E11 bad parameter, E12 syntax error), so the driver reads every setting back to confirm it.

  • FETCh? returns primary,secondary,bin (L/C/R/Z) or primary,bin (DCR); with tolerance on it is primary,deviation,bin in every mode, DCR included.

  • bin is the meter's last tolerance judgement, not a per-reading result: 1 pass, 0 fail, N none yet (or tolerance off). The meter only judges when a part is inserted while it is in local mode, and beeps once (pass) or three times (fail). While the driver polls in remote mode it neither beeps nor updates bin, so a part that drifts out of range keeps its old bin. For pass/fail per reading, compare deviation_pct with the range.

  • With the secondary display off, FUNC:IMPB? returns NULL and FETCh? still sends 0.0 for the secondary; the driver reports it as null.

  • Changing the frequency or level turns the secondary off (NULL), after which FETCh? keeps repeating the last secondary value instead of measuring it. The driver re-selects the secondary after those changes.

  • FUNC:IMPB? and FUNC:EQU? get no reply in DCR mode (the LCD shows E10), so they are never sent there; config reports frequency, level and equivalent as null in DCR, and setting them is refused before anything is sent.

  • Reopening the port immediately after closing it loses the next command (E10 on the LCD), so the driver keeps sessions 0.5 s apart and lets *GTL drain before closing.

  • macOS lets several programs open the same port, and they then read each other's replies. The driver takes an exclusive lock, so the CLI and Claude Desktop's server wait for each other (up to 10 s) instead of colliding.

  • Tolerance: turning CALC:TOL:STAT from OFF to ON captures the current reading as nominal and clears the range (so the range must be set after). While it is on, the secondary is NULL and FETCh?'s second field is the % deviation; turning it off leaves the secondary NULL. Changing the function over serial (even to DCR) leaves tolerance on with the old nominal, although the manual says a function change turns it off, so run tol on again after switching.

  • Recording (CALC:REC) covers the primary only (the secondary is always 0); the first MAX after switching on is left over from the previous run, and PRESent? always returns -----. The summarize stats are usually better.

  • Any command puts the meter in remote mode (RMT shown, buttons locked except USB). The manual calls that button the RMT key; on the meter it is labelled USB. Pressing it returns the meter to local until the next command, so during a log RMT only flashes off briefly. *LLO (local lockout), which the manual says also locks USB, has no visible effect on firmware 1.11, so the driver offers no panel lock.

  • The meter runs on its 9 V NiMH battery or a 12 V supply, not USB power; the USB cable only powers its CP2102. Left unused for the auto power-off delay (AoFF in the utility menu: 5, 15, 30 or 60 minutes or OFF; 5 here) it beeps for 10 seconds, then switches itself off; a button press during the beep resets the timer. It powers back up on C with no secondary, 1 kHz, 0.6 V, parallel, tolerance off, whatever it was set to before, and a powered-off meter does not answer over serial. Every serial command restarts the timer: a log with a reading every 10 s ran for 8 minutes, and the meter switched off 4 min 55 s after its last command. Auto-fetch output does not restart it: a stream stopped when the meter switched off 5 minutes after the last command (2026-11-09).

Tests

uv run pytest

Tests use a fake serial port and need no hardware. With a meter attached, uv run bk880 config is a quick end-to-end check.

License

MIT

Tool Schema Changelog

Recent tool additions, removals, and schema changes observed during successful MCP inspections.

Loading tool schema changelog…

    Related MCP Connectors

    Related MCP Servers

    • F
      license
      A
      quality
      C
      maintenance
      Enables LLMs like Claude to interact with PicoScope oscilloscopes for signal acquisition, measurement, and analysis. Supports device management, data capture, triggering, and signal generation through natural language commands.
      24
      8
      -
    • A
      license
      B
      quality
      D
      maintenance
      Enables agentic coding harnesses to control benchtop test equipment — waveform generator, multimeter, oscilloscope, and power supply — over SCPI, with typed tools and raw passthrough for investigating circuits.
      76
      1
      MIT
    • A
      license
      Not graded
      quality
      B
      maintenance
      Enables Claude or users to control SCPI-compatible oscilloscopes over Ethernet or USB, including measuring signals, changing settings, sending queries/writes, and capturing screenshots.
      MIT
    • A
      license
      Not graded
      quality
      B
      maintenance
      Enables AI assistants to automatically discover and control supported laboratory instruments over USB, LAN, and serial connections, including power-supply output, setpoint, and protection management plus multimeter measurements.
      4
      MIT