Sampling Frequency fS (kHz)
µC Clock Frequency (MHz)
Power Consumption (mW)
SNR
10.0
48
31.5
12.6
5.0
16
20.5
12.6
2.0
8
15.9
11.2
1.0
4
14.4
10.1
0.5
2
13.7
7.8
10
5
2
1
0.5
0
5
10
15
DSP Sampling Frequency fS (kHz)
SNR ()
Figure 17. SNR at different sampling frequencies. The EMG signal was measured at a muscle force at
20% of MVC and the noise was measured at a relaxed muscle. Note that the SNR is higher at 100%
MVC than at these measurements with lower muscle force.
4.9. Memory
The presented implementation including all configurations, filters, rectification and smoothing
and the decision algorithm requires 32.4% of the 256 kB program memory and 35.3% of the 32 kB
data memory.
4.10. Signal Delay