covnmr_sst
USP <761> system suitability: linewidth (FWHM) of the reference line,
signal-to-noise, peak asymmetry. Also the quick quality check: it absorbs
the retired covnmr_quality, whose S/N block is now the quality field.
`verdict` is pass / fail / incomplete, or "reference_not_found" when no
line is in reach of `reference_peak_ppm` (default 0.0, TMS). S/N is
evaluated only when `signal_ppm` is given, and a suite with an unevaluated
criterion reports "incomplete" (`not_tested`), never a pass. Limits default
to `max_linewidth_hz` 1.0 and `min_sn` 250. The linewidth is the
instrument's (no line broadening unless `line_broadening_hz` is given); the
spectrometer frequency comes from the file unless `field_mhz` is given.
Tool call template
Replace each placeholder with a value matching the input schema. Only required parameters are included; add optional parameters as needed.
{
"id": 1,
"jsonrpc": "2.0",
"method": "tools/call",
"params": {
"arguments": {
"path": "<path>"
},
"name": "covnmr_sst"
}
}Input Schema
| Name | Required | Description | Default |
|---|---|---|---|
| path | Yes | ||
| min_sn | No | ||
| api_key | No | ||
| nucleus | No | ||
| solvent | No | CDCl3 | |
| field_mhz | No | ||
| signal_ppm | No | ||
| max_linewidth_hz | No | ||
| line_broadening_hz | No | ||
| reference_peak_ppm | No |
Output Schema
| Name | Required | Description | Default |
|---|---|---|---|
| result | Yes |